Effect of Backscattered Radiation on X-Ray Image Contrast


  •  A. T. Naji    
  •  M. S. Jaafar    
  •  E. A. Ali    
  •  S. K. J. Al-Ani    

Abstract

This paper assesses the effect of backscattered radiation on X-ray image contrast by evaluating the effect of backscatter reduction on X-ray image contrast. Contrast test tool RMI Densitometer, and different types of fabricated anti backscattered grids have been utilized in this study. The measurements are recorded at different exposure parameters such as X-ray tube peak voltage (kVp), and X-rays intensities (mAs). For each exposure, the contrast of the image is evaluated by measuring the variation in optical densities for aluminium steps wedge. The results showed that the x-ray image contrast can be enhanced by decreasing the amount of backscattered radiation, also the fabricated anti backscattered grids have a remarkable effect in the improvement of X-ray image contrast according to grid’s capability in reducing backscattered radiation. In addition, the effectiveness of fabricated grids in improving image contrast depends on the grid’s material and the geometrical design, as well as the radiation exposure parameters. The image contrast enhancements increased up to 36% with the use of crossed iron steel grid which placed under the film screen combination during exposure.